Preventive Maintenance Instructions for the 3D-TOP Ceiling Stand, Siemens

Product group Radiological technology
Reference 03169/05

Siemens reminds all users of Systems with 3D-TOP Ceiling Stand about the importance of following the annual Preventive Maintenance Instructions, because it became aware that screws on the rollers may loosen when preventive maintenance instructions are not followed. This may cause one or more of the rollers to detach from the ceiling rails and presents a possible safety risk.